Repair a Welch 2025 or 2026 or 2027 diaphragm pump, "Dry-fast"


TYPICAL FAILURES:

  1. Excess noise. The bulk rubber below one diaphragm has fractured.
  2. Poor vacuum. One of two diaphragms has failed, or, oval-shaped internal rubber flap-valves too contaminated.
  3. Doesn't start (or starts after a delay.) Motor becomes hot. Bad start-capacitor, or destroyed diaphragm assembly is jamming the motor.
  4. Runs, but no vacuum. Hose leak, or totally failed diaphragm(s).


How to open up the stupid case!!! The top and rear are designed to slide backwards, once the six rear screws are removed. But this is almost impossible to do. Perhaps it worked fine, before all the steel panels were painted? Lots of prying finally frees up the top panel so it can slide backwards. Especially pry w/screwdriver at the top of the rear panel.





To inspect the diaphragms, first unscrew the hose connections to the two external connections to the pair of jars (unscrew inside the case.) Also unscrew the top of the vertical internal plastic hose near the rear of the unit (white poly right-angle hose connector.) Disconnect the thin rubber hose from the steel right-angle hose barb at the rear (it leads to the vac. regulator,) Finally remove the eight socket-cap screws from the corners of the two diaphragm blocks.



THE WHITE PLASTIC BLOCK FREQUENTLY ADHERES TO THE LIGHT-BLUE TEFLON DIAPHRAGM. I use flat screwdriver(s) to pry between the white plastic block and the thinner black plastic plate below. Perhaps carefully slide a blade above the blue-ish colored edge of the teflon diaphragm. Don't gouge anything, just pull the white block assembly upwards while prying all around the border, until it pops entirely loose. That exposes the circular teflon diaphragm. Repeat for the second white block. The center of each diaphragm should feel solid, not floppy like a bubble.




Another type of failure ...everything runs fine, but no vacuum shown on the meter. (Turn the "vacuum control" fully CW of course, and plug the vac inlet with a fingertip!) The leak was located in the small black hose leading to the pressure gauge, where hose-vibration eventually caused a crack. Cut off the hose at the crack, and it was still long enough to put on the meter-port. Warning: sometimes no leak exists, and the flap-valves inside the pump's top assembly are all crusty and need cleaning. Or, sometimes the expensive diaphragms have cracked from old age or from strong solvents. Then you'll need a Welch diaphragm rebuild kit, it's 2025K01 or 2025K02 for the 2025 model pump.)


Pump won't start? Or, long delay, while the motor heats up? Often the 15uF "motor-run" capacitor had decayed to 7uF, just barely enough to start the pump (after many, many seconds silent delay.) The motor coils overheat before the pump actually "gets over the hump" and starts spinning. (Carefully spin the front fan during power-up, and sometimes it will suddenly start.) If you don't have an exact replacement capacitor, just temporarily connect another few-uF capacitor in parallel with the existing one (using non-polarized, above-220VAC of course.) The pump will then run OK for a few months ...until the value of the original capacitor decreases even more.


Hint: search for ...run capacitor 16uf. Many are available if you search for 16uF rather than 15uF. Make sure it's 36mm diameter (under 1.4in,) otherwise it may not fit inside the pump housing. (Many of these similar capacitors are too wide, 2in diameter and 400VAC.)